    Modeling the Insertion of a Zickel Rod into the Femur

    Published: Jan 1989


      Format Pages Price  
    PDF (152K) 10 $25   ADD TO CART
    Complete Source PDF (5.5M) 10 $55   ADD TO CART


    Abstract

    A computer-aided design (CAD) system was used to model the insertion of a Zickel rod into a femur. Results show that although the shape of the rod and the intramedullary canal match closely in the fully inserted position, a considerable geometric mismatch can occur during rod insertion. The location of the starting hole and the valgus bend at the proximal end of the rod shaft can combine to create high stresses in the trochanteric region. High stresses can also occur during insertion between the canal and the distal posterior rod surface. These phenomena are the likely causes of the intraoperative complications of trochanteric fracture and the difficulty experienced in readjusting the anteversion once a rod has been inserted. Redesign of the distal end of the rod could reduce the difficulty in readjusting the anteversion.

    Keywords:

    intramedullary rods, Zickel rod, femur fractures, subtrochanteric fractures, geometric modeling, computer-aided design