ASTM E1657 - 98(2011)


ASTM E1657 - 98(2011) Standard Practice for Testing Variable-Wavelength Photometric Detectors Used in Liquid Chromatography


Active Standard ASTM E1657 Developed by Subcommittee: E13.19 |Book of Standards Volume: 03.06

more info 8 pages $ 41.00
more info 8 pages $ 41.00

Historical (view previous versions of standard) ASTM License Agreement Shipping & Handling

More E13.19 Standards Related Products
Copyright/Permissions Standard References

ASTM E1657

Significance and Use

Although it is possible to observe and measure each of the several characteristics of a detector under different and unique conditions, it is the intent of this practice that a complete set of detector specifications should be obtained under the same operating conditions. It should also be noted that to completely specify a detector's capability, its performance should be measured at several sets of conditions within the useful range of the detector. The terms and tests described in this practice are sufficiently general that they may be used regardless of the ultimate operating parameters.

Linearity and response time of the recorder or other readout device used should be such that they do not distort or otherwise interfere with the performance of the detector. This requires adjusting the gain, damping, and calibration in accordance with the manufacturer's directions. If additional electronic filters or amplifiers are used between the detector and the final readout device, their characteristics should also first be established.

1. Scope

1.1 This practice covers the testing of the performance of a variable-wavelength photometric detector (VWPD) used as the detection component of a liquid-chromatographic (LC) system operating at one or more wavelengths in the range 190 to 800 nm. Many of the measurements are made at 254 nm for consistency with Practice E685. Measurements at other wavelengths are optional.

1.2 This practice is intended to describe the performance of the detector both independently of the chromatographic system (static conditions) and with flowing solvent (dynamic conditions).

1.3 For general liquid chromatographic procedures, consult Refs (1-9).

1.4 For general information concerning the principles, construction, operation, and evaluation of liquid-chromatography detectors, see Refs (10, 11) in addition to the sections devoted to detectors in Refs (1-7).

1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.


2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.

ASTM Standards

E685 Practice for Testing Fixed-Wavelength Photometric Detectors Used in Liquid Chromatography



Keywords

linearity; noise measurement; photometric detector; variable wavelength; wavelength accuracy ; Drift; Linear changes/linearity; Liquid chromatography (LC); Long-term noise; Noise level measurement; Photometric detectors; Short-term noise; Variable-wavelength photometric detectors; VWPD (variable-wavelength photometric detectors); Wavelength accuracy



ICS Code

ICS Number Code 71.040.50 (Physicochemical methods of analysis)



DOI: 10.1520/E1657-98R11