ASTM WK94678
The test method requires the use of a 10-gram sample for analysis as specified in the sample preparation section. However, Note 6 gives an example which implies that a smaller sample size (e.g., approximately 5 mL) may be used, which is inconsistent with the method’s requirements. This revision corrects that inconsistency by reinforcing that only 10-gram samples are to be used. Additionally, while Note 6 refers to approximate volumes for convenience of estimating required masses, all internal standard additions, calibrations and sample preparation are performed by mass, as weighed on a balance. Commercial calibration standards are also provided in mass units. This revision clarifies the 10g sample requirement and aligns with mass for the available of calibration solutions.
Date Initiated: 04-22-2025
Technical Contact: Frank Di Sanzo