Work Item
ASTM WK81193

Revision of D1193-06(2018) Standard Specification for Reagent Water

Rationale

Auditors are requiring laboratories measure to ensure their water meets these specifications, however, the specifications for Type I and Type II water (particularly silica and TOC) have limits that are not readily measurable by the laboratory and there may be specific measurements that need change due to emerging contaminants, new improved sensitivity for measuring and applications where lower levels of contaminants requiring attention . Include clarification in the method, or revise the specification.

Details

Developed by Subcommittee: D19.02

Committee: D19

Staff Manager: Brian Milewski

Work Item Status

Date Initiated: 03-08-2022

Technical Contact: Gary Ganzi

Item: 001

Ballot: D19 (23-04)

Status: Withdrawn From Balloting