SYMPOSIA PAPER Published: 01 January 2004
STP11918S

Generalized Size and Shape Description of UHMWPE Wear Debris — A Comparison of Cross-Linked, Enhanced Fused, and Standard Polyethylene Particles

Source

Released wear debris of implants causes local inflammation of the host tissue if it is in a phagocytosable size. The purpose of this study was to compare particle size, shape, and number of three different types of UHMWPE. After wear testing, particles were isolated from the serum and analyzed using SEM. The parameters ‘equivalent circle diameter’ (ECD) and ‘equivalent shape ratio’ (ESR) were determined. Most of the generated debris was sub-micron in size. Classifying the particles into size groups demonstrated a non-linear correlation between size and shape for all three types of polyethylene: small particles were more round, large particles were more elongated. Based on this relationship, the generated number of particles and their total surface area were estimated and compared with calculations based on size alone.

Author Information

Sprecher, CM
AO-Research Institute, Davos, GR, Switzerland
Schneider, E
AO-Research Institute, Davos, GR, Switzerland
Wimmer, MA
Rush Presbyterian St. Luke's Medical Center, Chicago, IL
Price: $25.00
Contact Sales
Related
Reprints and Permissions
Reprints and copyright permissions can be requested through the
Copyright Clearance Center
Details
Developed by Committee: F04
Pages: 3–15
DOI: 10.1520/STP11918S
ISBN-EB: 978-0-8031-5492-6
ISBN-13: 978-0-8031-3474-4