Work Item
ASTM WK92508

Withdrawal of D6420-18 Standard Test Method for Determination of Gaseous Organic Compounds by Direct Interface Gas Chromatography-Mass Spectrometry

Rationale

Listing the chemicals in the scope is an odd choice. I would move this to summary (section 4) Negative: The scope should list the type of MS described. Is it EI? CI? TOF? The description in 1.4 seems very instrument specific, which is fine, but then the type of instrument needs to be declared. Table 1 AR, ARF, ARRF are not defined. should be in a footnote Table 5 SBP is a trade name. It should be Bonded; poly(dimethyl siloxane) phase 13 While there is a precision and bias statement, I think some of the data should be presented and not in a report that is not searchable via google.

Details

Developed by Subcommittee: D22.03

Committee: D22

Staff Manager: Ashley Wiand

Work Item Status

Date Initiated: 10-14-2024

Technical Contact: Dustin Poppendieck

Item: 000

Ballot:

Status: