Standard Guide for Detection and Quantification of Cleaning Markers (Analytes) for the Validation of Cleaning Methods for Reusable Medical Devices

Standard Guide for Detection and Quantification of Cleaning Markers (Analytes) for the Validation of Cleaning Methods for Reusable Medical Devices F3438-24 ASTM|F3438-24|en-US Standard Guide for Detection and Quantification of Cleaning Markers (Analytes) for the Validation of Cleaning Methods for Reusable Medical Devices Standard new BOS Vol. 13.02 Committee F04
$ 69.00 In stock
ASTM International

Significance and Use

5.1 This standard guide may be used by medical device manufacturers as part of their design plan and implementation of the validation of the cleaning instructions of their reusable medical devices.

5.2 This guide helps medical device manufacturers to identify the appropriate method(s) for detecting and quantifying markers for the simulated-use test soil (see Guide F3208), thereby evaluating whether the medical device can be adequately cleaned.

5.3 This guide describes various test methods for the different analytes.

5.4 This guide specifies the validation criteria for analyte detection methods.

Scope

1.1 This standard guide provides methods and considerations for detecting and quantifying test soil(s) from reusable medical device(s) that result from simulated-use testing of medical devices during validation of the cleaning procedures as described in the instructions for use (IFU) provided by the medical device manufacturer.

1.2 The methods described are for detecting and measuring markers (analytes) that are components within the most common test soils and are relevant to the clinical use of the device. Appropriate test soils without protein, carbon, or carbohydrates (for example, bone) will require other methods.

1.3 This is a part of a series of ASTM standard guides for validating cleaning instructions. The scope of the first guide in the series is selecting appropriate test soils (Guide F3208). The second in the series (Guide F3293) describes methods for inoculating medical devices with test soil. The third in the series (Guide F3321) describes methods for extracting soils for measuring residual soil on medical devices after the performance of cleaning process. This is the fourth guide in the series and describes the methods of detecting and quantifying residual analytes on the device.

1.4 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.