Significance and Use
5.1 Small quantities of sodium, 1 to 10 μg/L, can be significant in high pressure boiler systems and in nuclear power systems. Steam condensate from such systems must have less than 10 μg/L. In addition, condensate polishing system effluents should have less than 1 μg/L. Graphite furnace atomic absorption spectroscopy (GFAAS) represents technique for determining low concentrations of sodium.
1. Scope
1.1 This test method covers the determination of trace sodium in high purity water. The method range of concentration is 1 to 40 μg/L sodium. The analyst may extend the range once its applicability has been ascertained.
Note 1: It is necessary to perform replicate analysis and take an average to accurately determine values at the lower end of the stated range.
1.2 This test method is a graphite furnace atomic absorption spectrophotometric method for the determination of trace sodium impurities in ultra high purity water.
1.3 This test method has been used successfully with a high purity water matrix.2 It is the responsibility of the analyst to determine the suitability of the test method for other matrices.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
ASTM Standards
D1066 Practice for Sampling Steam
D1129 Terminology Relating to Water
D1192 Guide for Equipment for Sampling Water and Steam in Closed Conduits
D1193 Specification for Reagent Water
D2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
D3370 Practices for Sampling Water from Closed Conduits
D3919 Practice for Measuring Trace Elements in Water by Graphite Furnace Atomic Absorption Spectrophotometry
D4453 Practice for Handling of High Purity Water Samples
D5810 Guide for Spiking into Aqueous Samples
D5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis
ICS Code
ICS Number Code 71.040.50 (Physicochemical methods of analysis)
UNSPSC Code
UNSPSC Code
Link Here | |||
Link to Active (This link will always route to the current Active version of the standard.) | |||
DOI: 10.1520/D6071-14
Citation Format
ASTM D6071-14, Standard Test Method for Low Level Sodium in High Purity Water by Graphite Furnace Atomic Absorption Spectroscopy, ASTM International, West Conshohocken, PA, 2014, www.astm.org
Back to Top