ASTM D5085-21

Standard Test Method for Determination of Chloride, Nitrate, and Sulfate in Atmospheric Wet Deposition by Suppressed Ion Chromatography

Standard Test Method for Determination of Chloride, Nitrate, and Sulfate in Atmospheric Wet Deposition by Suppressed Ion Chromatography D5085-21 ASTM|D5085-21|en-US Standard Test Method for Determination of Chloride, Nitrate, and Sulfate in Atmospheric Wet Deposition by Suppressed Ion Chromatography Standard new BOS Vol. 11.07 Committee D22
$ 63.00 In stock

Significance and Use

5.1 This test method is for the determination of the anions: chloride, nitrate, and sulfate in atmospheric wet deposition.

5.2 Fig. X1.1 in the appendix represents cumulative frequency percentile concentration plots of chloride, nitrate, and sulfate obtained from analyses of over 5000 wet deposition samples. These data may be used as an aid in the selection of appropriate calibration solutions (2).

Scope

1.1 This test method is applicable to the determination of chloride, nitrate, and sulfate in atmospheric wet deposition samples (rain, snow, sleet, and hail) by suppressed ion chromatography. For additional applications, see to Test Method D4327.

1.2 The concentration ranges for this test method are as listed below. The range tested was confirmed using the interlaboratory collaborative test (see Table 1 for statistical summary of the collaborative test).

 

Method
Detection
L (mg/L) (1)

Range of
Method
(mg/L)

Range
Tested
(mg/L)

Chloride

0.03

0.09–2.0

0.15–1.36

Nitrate

0.03

0.09–5.0

0.15–4.92

Sulfate

0.03

0.09–8.0

0.15–6.52

1.3 The method detection limit (MDL) is based on single operator precision (1)2 and may be higher or lower for other operators and laboratories. The precision and bias data presented are insufficient to justify use at this low level; however, it has been reported that this test method is reliable at lower levels than those that were tested. The MDLs listed above were determined following the guidance in 40 CFR Part 136 Appendix B. Other approaches to the determination of MDLs may yield different MDLs.

1.4 Method Detection Limits will vary depending on the type and length of column(s) used, the composition and strength of eluent used, the bore size of the instrumentation (that is, microbore or standard bore), eluent flow rate and other variables between instruments. The method detection limits listed above are those used in determining the Precision and Bias of this method as given in Table 1.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 9.

1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.