    Rationale

    Under 9.1.2 it is written that: An instrument calibration is best carried out using an empty, clean, light-weight sample container Based on experience and knowledge of all critical parameters, this is sufficient to ensure correct results. I would like to propose to change the original statement above by the following: 9.1.2 An instrument calibration is best carried out using an empty and clean sample container of the same type (same dimensions, wall thickness, and material of construction) as that intended to be used for the measurements with samples.

    Citing ASTM Standards
    Back to Top
    Work Item Status

    Date Initiated:
    05-14-2020

    Technical Contact:
    Richard Turcotte

    Item:
    001

    Ballot:
    E27.02 (21-01)

    Status:
    Negative Votes Need Resolution