Standard Historical Last Updated: Feb 11, 2013 Track Document
ASTM D7353-07

Standard Practice for Sampling of Liquids in Waste Management Activities Using a Peristaltic Pump

Standard Practice for Sampling of Liquids in Waste Management Activities Using a Peristaltic Pump D7353-07 ASTM|D7353-07|en-US Standard Practice for Sampling of Liquids in Waste Management Activities Using a Peristaltic Pump Standard new BOS Vol. 11.04 Committee D34
$ 66.00 In stock

Significance and Use

This practice can be used in sampling drums, tanks, and similar containers and in sampling monitoring and waste wells including small diameter (1 in.) wells. The pump can collect samples from multiple depths. The samples can be high-viscosity fluids, aggressive and corrosive fluids, high-purity solutions and abrasive fluids. The pump can be used to mix samples (see D 6063).

Peristaltic pumps use a vacuum to transport the samples. This vacuum may cause some degassing and loss of volatile organic compounds (VOCs) from the sample. When precise quantitative data for VOCs and dissolved gases are not required, peristaltic pumps may be used.

The pump is self-priming, runs dry without damage, and is completely isolated from the pumped fluid. A sample can be taken on the intake or discharge side of the pump.

Some additional advantages of the peristaltic pump are: decontamination of the pump motor is not necessary and the tubing in the pump is disposable and easy to replace. The pumps can be easily started and stopped and can pump fluids at a wide range of pressures and flow rates.

The place, quality and quantity, frequency, and time of sampling is dependent upon the decisions that are to be made (see D 6250), sampling design (see D 6311), the sample, the heterogeneity of the samples (see D 5956), how representative the sample is (see D 6044), and the parameters to be tested as determined by the data quality objectives (DQOs) (see D 5792).

Scope

1.1 This practice covers the use of a peristaltic pump for sampling liquids from multiple depths. It is applicable for a wide range of fluids including: high-viscosity fluids, aggressive and corrosive fluids, high-purity solutions and abrasive fluids. It is especially useful for sampling liquids that require complete isolation from the pump.

1.2 This practice includes the determination of sample depth, pump set up and collecting a sample to be analyzed.

1.3 This practice is not intended to give detailed instructions for running a peristaltic pump or recommend which peristaltic pump to purchase. It instructs the field personnel how to connect the pump and collect a sample.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Price:
Contact Sales
Related
Reprints and Permissions
Reprints and copyright permissions can be requested through the
Copyright Clearance Center
Details
Book of Standards Volume: 11.04
Developed by Subcommittee: D34.01.03
Pages: 4
DOI: 10.1520/D7353-07
ICS Code: 23.080