ASTM D4599-03(2009)e1

Standard Practice for Measuring the Concentration of Toxic Gases or Vapors Using Length-of-Stain Dosimeters

Standard Practice for Measuring the Concentration of Toxic Gases or Vapors Using Length-of-Stain Dosimeters D4599-03R09E01 ASTM|D4599-03R09E01|en-US Standard Practice for Measuring the Concentration of Toxic Gases or Vapors Using Length-of-Stain Dosimeters Standard new BOS Vol. 11.07 Committee D22
$ 66.00 In stock

Significance and Use

The U.S. Occupational Safety and Health Administration (OSHA) in 29 CFR 1910.1000 Subpart Z designates that certain gases and vapors present in work place atmospheres must be controlled so that their concentrations do not exceed specified limits. Other countries have similar regulations.

This practice will provide a means for the determination of airborne concentrations of certain gases and vapors listed in 29 CFR 1910.1000 and in other countries regulations.

A partial list of chemicals for which this practice is applicable is presented in Annex A1 with current Threshold Limit Values (TLV) (2) and typical measurement ranges for the selected chemicals as obtained from various manufacturer's specifications.

This practice may be used for either personal or area monitoring.

Scope

1.1 This practice describes the detection and measurement of time weighted average (TWA) concentrations of toxic gases or vapors using length-of-stain colorimetric dosimeter tubes. A list of some of the gases and vapors that can be detected by this practice is provided in Annex A1. This list is given as a guide and should be considered neither absolute nor complete.

1.2 Length-of-stain colorimetric dosimeters work by diffusional sampling. The results are immediately available by visual observation; thus no auxiliary sampling, test nor analysis equipment are needed. The dosimeters, therefore, are extremely simple to use and very cost effective.

1.3 The values stated in SI units shall be regarded as the standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.