ASTM F2978 - 13

    Guide to Optimize Scan Sequences for Clinical Diagnostic Evaluation of Metal-on-Metal Hip Arthroplasty Devices using Magnetic Resonance Imaging

    Active Standard ASTM F2978 | Developed by Subcommittee: F04.22

    Book of Standards Volume: 13.02


      Format Pages Price  
    PDF Version 9 $42.00   ADD TO CART
    Print Version 9 $42.00   ADD TO CART


    Significance and Use

    5.1 Magnetic resonance imaging is ideally suited to image MOM hip arthroplasty due to its superior soft tissue contrast, multiplanar capabilities and lack of ionizing radiation. MR imaging is the most accurate imaging modality for the assessment of peri-prosthetic osteolysis and wear-induced synovitis (17-19).

    5.2 Before scanning a patient with a specific implant, the MR practitioner shall confirm that the device is MR Conditional and that the scan protocol to be used satisfies the conditions for safe scanning for the specific implant.

    5.3 This guide can be used to identify the following adverse events.

    5.3.1 Osteolysis—Magnetic resonance imaging is superior to conventional radiographs and CT in the assessment of peri-prosthetic osteolysis and has been shown to be the most accurate method to locate and quantify the extent of peri-prosthetic osteolysis (17, 18). On MR imaging, osteolysis appears as well marginated intraosseous intermediate to slightly increased signal intensity lesions that contrast with the high signal intensity of the intramedullary fat. A characteristic line of low signal intensity surrounds the area of focal marrow replacement, distinguishing the appearance of osteolysis from tumoral replacement of bone or infection (20).

    5.3.2 Component Loosening—While the data are preliminary, MR imaging can identify circumferential bone resorption that may indicate component loosening. Loosening may result from osteolysis, circumferential fibrous membrane formation or poor osseous integration of a non-cemented component. On MR imaging, component loosening typically manifests as circumferential increased signal intensity at the metallic-bone or cement-bone interface on fat-suppressed techniques (19). The finding of circumferential fibrous membrane formation or osteolysis also indicates potential loosening; this is in contrast to a well-fixed component, with high signal intensity fatty marrow directly opposed to the implant interface.

    5.3.3 Wear-Induced Synovitis—Magnetic resonance imaging is the most useful imaging modality to assess the intracapsular burden of wear-induced synovitis surrounding MOM arthroplasty (21). Preliminary data indicate that the signal characteristics of the synovial response on MR imaging correlate with the type of wear-induced synovitis demonstrated on histology at revision surgery (22). Low signal intensity debris is suggestive of metallic debris on histology. Mixed intermediate and low signal debris correlates with the presence of mixed polymeric (polyethylene and/or polymethyl methacrylate) and metallic debris at histology. Magnetic resonance imaging can demonstrate decompression of synovitis or fluid into adjacent bursae, such as the iliopsoas or trochanteric bursa, which can present as soft tissue masses or with secondary nerve compression. On occasion, wear-induced synovitis can result in a chronic indolent pattern of erosion of the surrounding bone, even in the absence of focal osteolytic lesions (6).

    1. Scope

    1.1 This guide describes the recommended protocol for magnetic resonance imaging (MRI) studies of patients implanted with metal-on-metal (MOM) devices to determine if the periprosthetic tissues are likely to be associated with an adverse local tissue reaction (ALTR). Before scanning a patient with a specific implant, the MR practitioner shall confirm that the device is MR Conditional and that the scan protocol to be used satisfies the conditions for safe scanning for the specific implant. This guide assumes that the MRI protocol will be applied to MOM devices while they are implanted inside the body. It is also expected that standardized MRI safety measures will be followed during the performance of this scan protocol.

    1.2 This guide covers the clinical evaluation of the tissues surrounding MOM hip replacement devices in patients using MRI. This guide is applicable to both total and resurfacing MOM hip systems.

    1.3 The protocol contained in this guide applies to whole body magnetic resonance equipment, as defined in section 2.2.103 of IEC 60601-2-33, Ed. 3.0, with a whole body radiofrequency (RF) transmit coil as defined in section 2.2.100. The RF coil should have quadrature excitation.

    1.4 The values stated in SI units are to be regarded as standard.

    1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. The user may consider all precautions and warnings provided in the MR system and hip implant labeling prior to determining the applicability of these protocols.


    2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.

    ASTM Standards

    A340 Terminology of Symbols and Definitions Relating to Magnetic Testing

    F2503 Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment

    IEC Standard

    IEC 60601-2-33, Ed. Medical Electrical Equipment--Part 2: Particular Requirements for the Safety of Magnetic Resonance Equipment for Medical Diagnosis, 2010


    ICS Code

    ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)

    ASTM F2978

    Citing ASTM Standards
    Back to Top