ASTM F2624 - 12

    Standard Test Method for Static, Dynamic, and Wear Assessment of Extra-Discal Single Level Spinal Constructs

    Active Standard ASTM F2624 | Developed by Subcommittee: F04.25

    Book of Standards Volume: 13.02


      Format Pages Price  
    PDF 15 $49.00   ADD TO CART
    Hardcopy (shipping and handling) 15 $49.00   ADD TO CART
    Standard + Redline PDF Bundle 30 $58.80   ADD TO CART


    Abstract

    This test method deals with static, dynamic, and wear testing of extra-discal motion preserving implants. These implants are intended to augment spinal stability without significant tissue removal while allowing motion of the functional spinal unit(s). Wear is assessed using a weight loss method and a dimensional analysis for determining wear of components used in extra-discal spinal motion preserving procedures, using testing medium as defined in this test method. This test method is not intended to address facet arthroplasty devices and any potential failure mode as it relates to the fixation of the device to its bony interfaces; and does not prescribe methods for assessing the mechanical characteristics of the device in translation. The static test includes the static flexion test, static extension test, static torsion test, static lateral bending test, and fatigue tests. Wear test includes flexion/extension wear assessment, rotational wear assessment, and bending wear assessment. The apparatus which shall be used includes implant components and spinal testing apparatus. The calculation and interpretation of wear results are also elaborated.

    This abstract is a brief summary of the referenced standard. It is informational only and not an official part of the standard; the full text of the standard itself must be referred to for its use and application. ASTM does not give any warranty express or implied or make any representation that the contents of this abstract are accurate, complete or up to date.

    Significance and Use

    This test method is designed to quantify the static and dynamic characteristics of different designs of single level spinal constructs. Wear may also be assessed for implants that allow motion using testing medium (see 6.1) for simulating the physiologic environment at 37°C. Wear is assessed using a weight loss method in addition to dimensional analyses. Weight loss is determined after subjecting the implants to dynamic profiles specified in this test method. This information will allow the manufacturer or end user of the product to understand how the specific device in question performs under the test conditions prescribed in this test method.

    This test method is intended to be applicable for single level extra-discal spinal constructs. Three different types of fixtures are specified for testing single level extra-discal spinal constructs (See Fig. 2, Fig. 4, and Fig. 5). See also Table 1.

    Implants may be designed using a variety of materials (for example, ceramics, metals, polymers, or combinations thereof), and it is the goal of this test method to enable a comparison of the static, dynamic, and wear properties generated by these devices, regardless of material and type of device.

    TABLE 1 Loading Modes and Associated Apparatus Listing Possible Tests That May Be Conducted (see 1.9)

    Note 1—For all loading modes, static, dynamic, and wear tests are described in this test method.

    Note 2—Offset refers to 8 mm of offset induced in the spinal construct (see Fig. 6) before subjecting the construct to rotational flexion/extension moments (see Fig. 2).

    Associated
    Apparatus
    Associated
    Figure
    Loading Mode
    RotationalFig. 2Flexion
    Extension
    Lateral Bending
    Axial Rotation
    Fig. 2 and Fig. 6Offset Flexion and Offset Extension
    ShearFig. 4Anterior/Posterior Shear
    Compression BendingFig. 5Compression Bending


    FIG. 4 Schematic of Anterior/Posterior Shear Testing Apparatus


    FIG. 5 Schematic of Single Level Compression Bending Test

    1. Scope

    1.1 This test method describes methods to assess the static and dynamic properties of single level spinal constructs.

    1.2 An option for assessing wear using a weight loss method and a dimensional analysis is given. This method, described herein, is used for the analysis of devices intended for motion preservation, using testing medium as defined in this standard (6.1).

    1.3 This test method is not intended to address any potential failure mode as it relates to the fixation of the device to its bony interfaces.

    1.4 It is the intent of this test method to enable single level extra-discal spinal constructs with regard to kinematic, functional, and wear characteristics when tested under the specified conditions.

    1.5 This test method is not intended to address facet arthroplasty devices.

    1.6 In order that the data be reproducible and comparable within and between laboratories, it is essential that uniform procedures be established. This test method is intended to facilitate uniform testing methods and data reporting.

    1.7 The motion profiles specified by this test method do not necessarily accurately reproduce those occurring in vivo. Rather this method provides useful boundary/endpoint conditions for evaluating implant designs in a functional manner.

    1.8 This test method is not intended to be a performance standard. It is the responsibility of the user of this test method to characterize the safety and effectiveness of the device under evaluation.

    1.9 Multiple test methods are included in this standard. However, it must be noted that the user is not obligated to test using all of the described methods. Instead, the user should only select test methods that are appropriate for a particular device design. In most instances, only a subset of the herein described test methods will be required.

    1.10 The values stated in SI units are to be regarded as the standard with the exception of angular measurements, which may be reported in either degrees or radians. No other units of measurement are included in this standard.

    1.11 This test method does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this test method to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.


    2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.

    ASTM Standards

    E2309 Practices for Verification of Displacement Measuring Systems and Devices Used in Material Testing Machines

    F561 Practice for Retrieval and Analysis of Medical Devices, and Associated Tissues and Fluids

    F1714 Guide for Gravimetric Wear Assessment of Prosthetic Hip Designs in Simulator Devices

    F1717 Test Methods for Spinal Implant Constructs in a Vertebrectomy Model

    F1877 Practice for Characterization of Particles

    F2003 Practice for Accelerated Aging of Ultra-High Molecular Weight Polyethylene after Gamma Irradiation in Air

    F2423 Guide for Functional, Kinematic, and Wear Assessment of Total Disc Prostheses


    ICS Code

    ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)

    UNSPSC Code

    UNSPSC Code 42295500(Surgical implants and expanders and extenders and surgical wires and related products)


    DOI: 10.1520/F2624-12