ASTM F2119-07(2013)

Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants (Withdrawn 2022)

Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants (Withdrawn 2022) F2119-07R13 ASTM|F2119-07R13|en-US Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants (Withdrawn 2022) Standard new BOS Vol. 13.01 Committee F04
$ 0.00 Out of stock

Significance and Use

5.1 This test method provides a quantified measure of the image artifact produced under a standard set of scanning conditions.

5.2 This test method applies only to passive implants that have been established to be MR-Safe or MR-Conditional.

Scope

1.1 This test method characterizes the distortion and signal loss artifacts produced in a magnetic resonance (MR) image by a passive implant (implant that functions without the supply of electrical or external power). Anything not established to be MR-Safe or MR-Conditional is excluded.

1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

Language unavailable