Standard Active Last Updated: Jan 06, 2023 Track Document
ASTM E3029-15(2023)

Standard Practice for Determining Relative Spectral Correction Factors for Emission Signal of Fluorescence Spectrometers

Standard Practice for Determining Relative Spectral Correction Factors for Emission Signal of Fluorescence Spectrometers E3029-15R23 ASTM|E3029-15R23|en-US Standard Practice for Determining Relative Spectral Correction Factors for Emission Signal of Fluorescence Spectrometers Standard new BOS Vol. 03.06 Committee E13
$ 63.00 In stock

Significance and Use

3.1 Calibration of the responsivity of the detection system for emission (EM) as a function of EM wavelength (λEM), also referred to as spectral correction of emission, is necessary for successful quantification when intensity ratios at different EM wavelengths are being compared or when the true shape or peak maximum position of an EM spectrum needs to be known. Such calibration methods are given here and summarized in Table 1. This type of calibration is necessary because the spectral responsivity of a detection system can change significantly over its useful wavelength range (see Fig. 1). It is highly recommended that the wavelength accuracy (see Test Method E388) and the linear range of the detection system (see Guide E2719 and Test Method E578) be determined before spectral calibration is performed and that appropriate steps are taken to insure that all measured intensities during this calibration are within the linear range. For example, when using wide slit widths in the monochromators, attenuators may be needed to attenuate the excitation beam or emission, thereby, decreasing the fluorescence intensity at the detector. Also note that when using an EM polarizer, the spectral correction for emission is dependent on the polarizer setting. (2) It is important to use the same instrument settings for all of the calibration procedures mentioned here, as well as for subsequent sample measurements.

FIG. 1 Example of Relative Spectral Responsivity of Emission Detection System (Grating Monochromator-PMT Based), (see Test Method E578) for which a Correction Needs to be Applied to a Measured Instrument-Specific Emission Spectrum to Obtain its True Spectral Shape (Relative Intensities).

Example of Relative Spectral Responsivity of Emission Detection System (Grating Monochromator-PMT Based),(see Test Method ) for which a Correction Needs to be Applied to a Measured Instrument-Specific Emission Spectrumto Obtain its True Spectral Shape (Relative Intensities).Example of Relative Spectral Responsivity of Emission Detection System (Grating Monochromator-PMT Based),(see Test Method ) for which a Correction Needs to be Applied to a Measured Instrument-Specific Emission Spectrumto Obtain its True Spectral Shape (Relative Intensities).

3.2 When using CCD or diode array detectors with a spectrometer for λEM selection, the spectral correction factors are dependent on the grating position of the spectrometer. Therefore, the spectral correction profile versus λEM must be determined separately for each grating position used. (3)

3.3 Instrument manufacturers often provide an automated procedure and calculation for a spectral correction function for emission, or they may supply a correction that was determined at the factory. This correction can often be applied during spectral collection or as a post-collection correction. The user should be advised to verify that the automated vendor procedure and calculation or supplied correction are performed and determined according to the guidelines given within this standard.

Scope

1.1 This practice (1)2 describes three methods for determining the relative spectral correction factors for grating-based fluorescence spectrometers in the ultraviolet-visible spectral range. These methods are intended for instruments with a 0°/90° transmitting sample geometry. Each method uses different types of transfer standards, including 1) a calibrated light source (CS), 2) a calibrated detector (CD) and a calibrated diffuse reflector (CR), and 3) certified reference materials (CRMs). The wavelength region covered by the different methods ranges from 250 nm to 830 nm with some methods having a broader range than others. Extending these methods to the near infrared (NIR) beyond 830 nm will be discussed briefly, where appropriate. These methods were designed for scanning fluorescence spectrometers with a single channel detector, but can also be used with a multichannel detector, such as a diode array or a CCD.

1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Price:
Contact Sales
Related
Reprints and Permissions
Reprints and copyright permissions can be requested through the
Copyright Clearance Center
Details
Book of Standards Volume: 03.06
Developed by Subcommittee: E13.01
Pages: 5
DOI: 10.1520/E3029-15R23
ICS Code: 17.180.01