ASTM D7102 - 10

    Standard Guide for Determination of Endotoxin on Sterile Medical Gloves

    Active Standard ASTM D7102 | Developed by Subcommittee: D11.40

    Book of Standards Volume: 09.02


      Format Pages Price  
    PDF 5 $43.00   ADD TO CART
    Hardcopy (shipping and handling) 5 $43.00   ADD TO CART
    Standard + Redline PDF Bundle 10 $51.60   ADD TO CART


    Significance and Use

    This guide establishes a standard sample preparation method and provides a description of four established and recognized test methods for the determination of endotoxin on medical gloves. If interferences in a sample yield suspect results, a second method should be used.

    This guide is appropriate for testing final product that has been subjected to all processes that could influence the final endotoxin level (either microbial contamination or processing agents/raw materials contaminated with endotoxin). As raw materials and processing conditions vary from lot to lot with regard to these parameters, it is appropriate to test for endo-toxin on a routine basis if a product endotoxin claim is to be made (for example, non-pyrogenic). The user may find it beneficial to incorporate endotoxin testing for vulnerable areas of their manufacturing process as an alert mechanism.

    1. Scope

    1.1 This guide covers a selection of methodologies for the determination of bacterial endotoxin on gloves when such a determination is appropriate.

    1.2 As bacteria may continue to grow on non-sterile gloves, reportable endotoxin levels are only appropriate for gloves labeled as sterile. Because most environments contain endotoxin, once a box of gloves is opened and the gloves are manipulated, endotoxin levels will increase making it inappropriate to report endotoxin levels on boxed gloves (ex. examination gloves). This is true even if the box had undergone sterilization prior to distribution.

    1.3 This guide may also be appropriate for internal quality control or alert purposes at different stages of manufacturing or during process change evaluations.

    1.4 This guide is not applicable to the determination of pyrogens other than bacterial endotoxins.

    1.5 The sample preparation method described must be used regardless of the test method selected. This method does not describe laboratory test method validation, analyst qualification, or reagent confirmation. Product-specific validation is addressed.

    1.6 The safe and proper use of medical gloves is beyond the scope of this guide.

    1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.


    2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.

    ANSI Standard

    ANSI/AAMIST72:2002 Bacterial Endotoxins--Test Methodologies, Routine Monitoring and Alternatives to Batch Testing

    EN Standard

    EN455-3:1999 Medical Gloves for Single Use--Part 3: Requirements and Testing for Biological Evaluation


    ICS Code

    ICS Number Code 11.140 (Hospital equipment)

    UNSPSC Code

    UNSPSC Code 42132203(Medical exam or non surgical procedure gloves)


    Referencing This Standard

    DOI: 10.1520/D7102-10

    ASTM International is a member of CrossRef.

    Citation Format

    ASTM D7102-10, Standard Guide for Determination of Endotoxin on Sterile Medical Gloves, ASTM International, West Conshohocken, PA, 2010, www.astm.org