ASTM D6956 - 11

    Standard Guide for Demonstrating and Assessing Whether a Chemical Analytical Measurement System Provides Analytical Results Consistent with Their Intended Use

    Active Standard ASTM D6956 | Developed by Subcommittee: D34.01.01

    Book of Standards Volume: 11.04


    Buy Standard (PDF)
    12 pages
    $48.00
    Buy Standard (Print)
    12 pages
    $48.00
    Buy Standard + Redline (PDF)
    24 pages
    $57.60

    Historical (view previous versions of standard)


    ASTM D6956

    Significance and Use

    This guide is intended for use by both generators and users of analytical results. It is intended to promote consistent demonstration and documentation of the quality of the measurement results and facilitate determination of the validity of measurements for their intended use.

    This guide specifies documentation that a laboratory should supply with the analytical results to establish that the resulting measurements: (1) meet measurement quality requirements; (2) are suitable for their intended use; and (3) are technically defensible.

    While the guide describes information that the measurement results provider needs to give the user/decision maker, in order for measurement providers to supply data users with appropriate data, information is needed from the data user. Examples of information that the user should provide to the laboratory, in addition to the analytes of concern (including the form of the analyte that is to be determined, for example, total lead, dissolved lead, organic lead, inorganic lead), include but are not limited to:

    Type of material (that is, matrixfresh or salt water, coal fly ash, sandy loam soil, wastewater treatment sludge),

    Maximum sample holding time,

    Projected sampling date and delivery date to the laboratory,

    Method of chemical preservation (for example, not preserved, chemical used),

    Chain-of-custody requirements, if any,

    Analytical methods that must be used, if any,

    Measurement quality requirements expressed as DQOs or MQOs and action limits,

    Allowable interferences as described in 10.4,

    Documentation requirement, and

    Subcontracting restrictions/requirements.

    Users/decision makers should consult with the laboratory about these issues during the analytical design stage. This will allow the design of sample collection process and project schedule to accommodate the laboratory activities necessary to determine the desired level of measurement quality. The number of samples, budgets, and schedules should also be discussed.

    1. Scope

    1.1 This guide describes an approach for demonstrating the quality of analytical chemical measurement results from the application of a measurement system (that is, method or sequence of methods) to the analysis of environmental samples of soil, water, air, or waste. The purpose of such measurements can include demonstrating compliance with a regulatory limit, determining whether a site is contaminated above some specified level, or determining treatment process efficacy.

    1.2 This guide describes a procedure that can be used to assess a measurement system used to generate analytical results for a specific purpose. Users and reviewers of the analytical results can determine, with a known level of confidence, if they meet the quality requirements and are suitable for the intended use.

    1.3 This protocol does not address the general components of laboratory quality systems necessary to ensure the overall quality of laboratory operations. For such systems, the user is referred to International Standards Organization (ISO) Standard 17025 or the National Environmental Laboratory Accreditation Conference (NELAC) laboratory accreditation standards.

    1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

    1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.


    2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.

    ASTM Standards

    D4687 Guide for General Planning of Waste Sampling

    D5283 Practice for Generation of Environmental Data Related to Waste Management Activities: Quality Assurance and Quality Control Planning and Implementation

    D5681 Terminology for Waste and Waste Management

    D5792 Practice for Generation of Environmental Data Related to Waste Management Activities: Development of Data Quality Objectives

    D5956 Guide for Sampling Strategies for Heterogeneous Wastes

    D6044 Guide for Representative Sampling for Management of Waste and Contaminated Media

    D6233 Guide for Data Assessment for Environmental Waste Management Activities

    D6250 Practice for Derivation of Decision Point and Confidence Limit for Statistical Testing of Mean Concentration in Waste Management Decisions

    D6311 Guide for Generation of Environmental Data Related to Waste Management Activities: Selection and Optimization of Sampling Design

    D6582 Guide for Ranked Set Sampling: Efficient Estimation of a Mean Concentration in Environmental Sampling

    D6597 Practice for Assessment of Attaining Clean Up Level for Site Closure

    Other Documents

    QuantifyingUncertain


    ICS Code

    ICS Number Code 71.040.40 (Chemical analysis)

    UNSPSC Code

    UNSPSC Code 77100000(Environmental management)


    DOI: 10.1520/D6956-11