ASTM D5462 - 08 Standard Test Method for On-Line Measurement of Low-Level Dissolved Oxygen in Water


Active Standard ASTM D5462 Developed by Subcommittee: D19.03 |Book of Standards Volume: 11.02

more info 6 pages $ 41.00
more info 6 pages $ 41.00
why redline? 12 pages $ 49.20

Historical (view previous versions of standard) ASTM License Agreement Shipping & Handling

More D19.03 Standards Related Products
Copyright/Permissions Standard References

ASTM D5462

Significance and Use

DO may be either a corrosive or passivating agent in boiler/steam cycles and is therefore controlled to specific concentrations that are low relative to environmental and wastewater treatment samples. Out-of-specification DO concentrations may cause corrosion in boiler systems, which leads to corrosion fatigue and corrosion productsall detrimental to the life and efficient operation of a steam generator. The efficiency of DO removal from boiler feedwater by mechanical or chemical means, or both, may be monitored by continuously measuring the DO concentration before and after the removal process with on-line instrumentation. DO measurement is also a check for air leakage into the boiler water cycle.

Guidelines for feedwater to high-pressure boilers with all volatile treatment generally require a feedwater DO concentration below 5 μg/L (3).

Boiler feedwater with oxygenated treatment is maintained in a range of 50 to 300 μg/L DO (4).

In microelectronics production, DO can be detrimental in some manufacturing processes, for example, causing undesirable oxidation on silicon wafers.

1. Scope

1.1 This test method covers the on-line determination of dissolved oxygen (DO) in water samples primarily in ranges from 0 to 500 μg/L (ppb), although higher ranges may be used for calibration. On-line instrumentation is used for continuous measurements of DO in samples that are brought through sample lines and conditioned from high-temperature and high-pressure sources when necessary.

1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazards statements, see 6.5.


2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.

ASTM Standards

D1066 Practice for Sampling Steam

D1129 Terminology Relating to Water

D1193 Specification for Reagent Water

D2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water

D3370 Practices for Sampling Water from Closed Conduits

D3864 Guide for Continual On-Line Monitoring Systems for Water Analysis



Keywords

dissolved oxygen; galvanic cell; oxygen; polarographic sensor; Boiler feedwater; Dissolved oxygen (DO); Galvanic cell; Online monitoring (in water analysis); Oxygen demand (in water); Polarographic methods; Total oxygen demand (TOD);



ICS Code

ICS Number Code 13.060.50 (Examination of water for chemical substances)



DOI: 10.1520/D5462-08