Infrared Quantitative Evidential Breath-Alcohol Analyzers: In Vitro Accuracy and Precision Studies
Goldberger, BA Graduate student in forensic toxicology, University of Maryland, School of Medicine, Baltimore, MD.
Caplan, YH Professor and director of forensic toxicology, University of Maryland, School of Medicine, Baltimore, MD and toxicologist, Office of the Chief Medical Examiner, Baltimore, MD.
Abstract
The in vitro accuracy and precision of four infrared breath-alcohol analyzers, the Alcotest 7010, BAC Verifier, Intoxilyzer 5000, and Intoximeter 3000, were studied with a protocol adapted from portions of the U.S. Department of Transportation's (DOT) “Standard for Devices to Measure Breath Alcohol.” Statistical evaluation of these studies indicated that all instruments met or exceeded the performance requirements modified from the U.S. DOT Standard for quantitative evidential breath-alcohol analyzers.
Keywords:
breath-alcohol testing devices, forensic science, protocols, statistical analysis, toxicology
Paper ID: JFS311860016
ASTM International is a member of CrossRef.
Author Goldberger BA, Caplan YH
Title Infrared Quantitative Evidential Breath-Alcohol Analyzers: In Vitro Accuracy and Precision Studies
Symposium ,
Committee on
|