    Surface Temperature Sensor Calibration: In Situ Technique

    Published: Jan 1985


      Format Pages Price  
    PDF (160K) 7 $25   ADD TO CART
    Complete Source PDF (3.3M) 7 $55   ADD TO CART


    Abstract

    The surface temperature dectors on a guarded hot plate can be calibrated in place using differential thermopiles. A ten-junction Type T thermopile was used to calibrate a 1.22-m (4-ft) guarded hot plate using water triple point cells as the reference temperature. The stability and precision of these thermopiles, over a year's span, was on the order of 20 to 30 millidegrees.

    Keywords:

    guarded hot plate, thermopiles, senor calibration, thermopiles


    Author Information:

    Schumann, SP
    Engineer, Technical Center, Owens Corning Fiberglas, Granville, OH