STP789: Comments on Calibration and Design of a Heat Flow Meter

    Bomberg, M
    Research Officer, Division of Building Research, National Research Council of Canada, Ottawa, Ont.

    Solvason, KR
    Research Officer, Division of Building Research, National Research Council of Canada, Ottawa, Ont.

    Pages: 16    Published: Jan 1983


    Abstract

    After a review of the calibration of two heat flow meter (HFM) transducers, the HFM has been redesigned. The new design produces a calibration independent of the characteristics of the material being tested; that is, it produces the same calibration curve for thick, low-density specimens as for thin, high-density ones. Calibration within ±1 percent of transfer standard accuracy is discussed; the HFM precision for thick specimens of low-density glass fiber appears to be higher than that of a guarded hot plate under the same testing conditions.

    Keywords:

    heat flow meter, guarded hot plate, thermal conductivity, thermal resistance, HFM calibration, HFM design, precision and accuracy