STP953

    Histological Comparison of Biological Fixation and Bone Modeling with Canine and Human Porous-Coated Hip Prostheses

    Published: Jan 1987


      Format Pages Price  
    PDF Version (916K) 22 $25   ADD TO CART
    Complete Source PDF (11M) 22 $82   ADD TO CART


    Abstract

    Undecalcified hard-section histology was performed on twelve canine and eleven human porous-coated femoral hip prostheses implanted without bone cement. The period of implantation varied from four weeks to seven years. All of the canine and nine of the eleven human implants showed regions of substantial bone ingrowth along the porous-coated stem region. Areas of the two human implants ingrown with fibrous tissue were characterized radiographically by a radiopaque line surrounding, but separated from, the implant surface. Resorptive bone modeling was observed with both implant types, particularly proximally. New bone formation and hypertrophy typically occur at regions of proximity to or contact with the endosteal cortex and, in the case of proximally porous coated implants, at the junction of the porous and smooth implant surfaces. The smooth-surfaced distal region of proximally coated implants is typically surrounded by a thin shell of bone that is separated from the implant surface. Overall, resorptive and formative bony change appears more exaggerated in the canine, but the radiographic and histologic pictures of canine and human implants bear resemblances.

    Keywords:

    porous implants, bone ingrowth, histology


    Author Information:

    Bobyn, JD
    Research director, Orthopaedic Research Laboratory, Montreal General Hospital Research Institute, McGill University, Montreal, Quebec

    Engh, CA
    Medical director, Anderson Orthopaedic Research Institute, Arlington, VA

    Pilliar, RM
    Professor of biomaterials, University of Toronto, Toronto, Ontario


    Paper ID: STP25231S

    Committee/Subcommittee: F04.93

    DOI: 10.1520/STP25231S