Digital Library / STP / STP1015-EB / STP18723S



Digital Fluence Profiling System and Software
Seitel, SC
Montana Laser Optics, Inc., MY

Rypma, RL
Montana Laser Optics, Inc., MY


Pages: 1    Published: Jan 1988


Download this paper for $25 PDF (15K)          View License Agreement
        Click here to download the complete source publication for $77 PDF (14M)


Source: STP1015-EB


Abstract

An advanced digital fluence profiling system based on a windowless GE TN2502 or TN2509 silicon array camera and Poynting Products 505RM frame grabber is described. The camera array is located precisely in the target plane or equivalent, and synchronized to the pulsed laser source with a flashlamp radiation sensor circuit. Data acquired by the frame grabber are transferred to an HP Series 200 computer for analysis.

Analysis is performed with BEAMCODE 2.0 software which permits near real time extraction of the peak fluence to total energy ratio in a single pulse to better than 20 percent. When corrected to camera nonuniformity and compared to large-aperture calorimetry, accuracy exceeds 10 percent. Other functions include measurement of pointing stability (beam wander) and divergence, fit-to-gaussian, and graphics data displays in contour and pseudo-3D formats.


Keywords:
computer control, laser beam profile, laser diagnostics, laser fluence measurement

Paper ID: STP18723S
Committee/Subcommittee: E13.03
DOI: 10.1520/STP18723S
CrossRef ASTM International is a member of CrossRef.