STP1272

    Wear of Titanium 6-4 Alloy in Laboratory Tests and in Retrieved Human Joint Replacements

    Published: Jan 1996


      Format Pages Price  
    PDF (516K) 28 $25   ADD TO CART
    Complete Source PDF (9.2M) 28 $87   ADD TO CART


    Abstract

    Wear of titanium 6-4 alloy against UHMW polyethylene was compared to that with cobalt chromium or stainless steel, in several types of laboratory wear machines and on retrieved failed hip prostheses. In the absence of third-body abrasive contaminants, titanium alloy tended to produce polyethylene wear equal to and sometimes lower than the other alloys. Severe, self-perpetuating abrasive-corrosive wear of the titanium alloy could be initiated by the entrapment of third-body particles between the metal and polymer, sometimes with particles of acrylic surgical cement and always with metallic particles, as from porous coated prostheses. The micro-scratching (dulling) commonly observed on the bearing surfaces of retrieved titanium alloy components appeared to be primarily due entrapment of third-body acrylic and/or metallic particles that were initially generated by fretting of loosened prostheses, for example, at the stem-cement interface. Nitrogen hardening of the titanium surfaces, e.g., by diffusion or ion-implanting, provided additional protection against cement particles, but did not prevent severe abrasion by entrapped metallic particles.

    Keywords:

    titanium 6-4 alloy, hip prosthesis, wear, nitriding, UHMW polyethylene, joint simulator


    Author Information:

    McKellop, HA
    J. Vernon Luck Orthopaedic Research Center, Los Angeles Orthopaedic HospitalUniversity of Southern California, Los Angeles, CA

    Rostlund, T
    University of Gothenbuurg, Gothenburg,

    Ebramzadeh, E
    J. Vernon Luck Orthopaedic Research Center, Los Angeles Orthopaedic HospitalUniversity of Southern California, Los Angeles, CA

    University of Gothenbuurg, Gothenburg,

    Sarmiento, A
    J. Vernon Luck Orthopaedic Research Center, Los Angeles Orthopaedic HospitalUniversity of Southern California, Los Angeles, CA


    Paper ID: STP16085S

    Committee/Subcommittee: F04.18

    DOI: 10.1520/STP16085S


    CrossRef ASTM International is a member of CrossRef.