STP1144

    Histopathological Effects of Ultrahigh-Molecular-Weight Polyethylene and Metal Wear Debris in Porous and Cemented Surface Replacements

    Published: Jan 1992


      Format Pages Price  
    PDF (636K) 14 $25   ADD TO CART
    Complete Source PDF (6.5M) 14 $65   ADD TO CART


    Abstract

    The replacement of cement with biological ingrowth systems for fixation has not resolved the problem of aseptic loosening nor that of osteolysis in association with metal and polyethylene wear debris. In order to investigate the role of wear debris in the failure of surface replacements, a histopathological study of ten titanium alloy porous-coated (PSR) and ten cobalt-chrome (Co-Cr) alloy cemented components was conducted. Wear-debris-laden histiocytes (metal and polyethylene) were associated with massive osteolysis in several of the PSR components, but bone loss in the cemented Co-Cr components was confined to localized areas adjacent to the cement membrane. The results of this study suggest that histiocytes activated by the ingestion of fine wear debris are responsible for bone loss. The mechanism of bone loss shown by this “model” is applicable to all joint replacement implant systems where wear debris is generated.

    Keywords:

    particulates, implants, arthroplasty, wear debris, histiocytes, titanium, polyethylene


    Author Information:

    Campbell, P
    Director, Implant Retrieval Laboratory, Joint Replacement Institute, Los Angeles, CA

    Nasser, S
    Associate professor, Emory University School of Medicine, Atlanta, GA

    Kossovsky, N
    Assistant professor, UCLA School of Medicine, Los Angeles, CA

    Amstutz, HC
    Medical director, Joint Replacement Institute, Los Angeles, CA


    Paper ID: STP14850S

    Committee/Subcommittee: F04.22

    DOI: 10.1520/STP14850S


    CrossRef ASTM International is a member of CrossRef.