Extrusion of Interbody Fusion DevicesClinical Examples
Theiss SM


Pages: 5    Published: Jan 2003


Download this paper for $25 PDF (112K)          View License Agreement
        Click here to download the complete source publication for $118 PDF (5.0M)


Source: STP1431-EB


Abstract
Extrusion of lumbar interbody fusion devices is a rare, though well-described, complication. Extrusion has been seen with a variety of interbody devices and surgical techniques. It is most commonly seen when implants are placed using a posterior approach. Extrusion results from instability of the final construct following implantation, or misplacement of the device. Instability can result from placement of an undersized implant or excessive resection of the facet joints. In each case, an error in surgical technique can be identified which leads to the extrusion. A design flaw of the implant is generally not the cause for the extrusion.


Keywords:
interbody fusion, fusion cage, extrusion, lumbar, PLIF