A Round Robin Interlaboratory Comparison of Thermal Conductivity Testing Using the Guarded Hot Plate up to 1000°C Pages: 15 Published: Jan 2002
Download this paper for $25
PDF (248K)
View License Agreement Source: STP1426-EB Abstract A round robin interlaboratory comparison of thermal conductivity measurements was performed using the Guarded Hot Plate apparatus at temperatures from about 0°C to 1000 °C. There were twelve participating laboratories including four national laboratories in three different countries. A statistical analysis was performed and the variation in measurement results is discussed. Both within laboratory and between laboratory variability is analyzed. A definite conclusion of this comparison is that measurement variation increases progressively with increasing temperature. As a result there is a need for the U.S. national laboratory, the National Institute of Standards & Technology, to develop high temperature testing capability and then high temperature thermal conductivity reference standards. Keywords: Interlaboratory comparison, round robin, thermal conductivity, thermal measurements, Guarded Hot Plate, C 177, ISO-8302, high temperature, C 1114 Paper ID: STP11008S Committee/Subcommittee: C16.30 DOI: 10.1520/STP11008S ASTM International is a member of CrossRef. | ||