    A Round Robin Interlaboratory Comparison of Thermal Conductivity Testing Using the Guarded Hot Plate up to 1000°C

    Published: Jan 2002


      Format Pages Price  
    PDF Version (240K) 15 $25   ADD TO CART
    Complete Source PDF (7.6M) 15 $143   ADD TO CART


    Abstract

    A round robin interlaboratory comparison of thermal conductivity measurements was performed using the Guarded Hot Plate apparatus at temperatures from about 0°C to 1000 °C. There were twelve participating laboratories including four national laboratories in three different countries. A statistical analysis was performed and the variation in measurement results is discussed. Both within laboratory and between laboratory variability is analyzed. A definite conclusion of this comparison is that measurement variation increases progressively with increasing temperature. As a result there is a need for the U.S. national laboratory, the National Institute of Standards & Technology, to develop high temperature testing capability and then high temperature thermal conductivity reference standards.

    Keywords:

    Interlaboratory comparison, round robin, thermal conductivity, thermal measurements, Guarded Hot Plate, C 177, ISO-8302, high temperature, C 1114