Volume 33, Issue 5 (September 2005)

    Thermal Control of the Test Liquid in Vibratory Cavitation Erosion Tests

    (Received 19 May 2004; accepted 1 September 2005)

    Published Online: 2005

    CODEN: JTEOAD

      Format Pages Price  
    PDF 9 $25   ADD TO CART


    Abstract

    Implosion of vaporous cavities is exothermal, thereby heating the cavitating liquid. In vibratory cavitation erosion (VCE) tests the erosion rate increases with test liquid temperature. In order to obtain the erosive properties of a given material at a given temperature, it is necessary to control the test liquid temperature. The prevailing ASTM standard specifies cooling of the test liquid, and allows a maximal temperature rise of 4°C during the VCE test. However, cooling alone does not ensure the required uniformity of test liquid temperature near the specimen.

    In order to obtain a uniform test liquid temperature near the specimen during VCE tests, a thermal control system is recommended. This system consists of three subsystems for cooling, heating, and mixing the test liquid. Correct location of the temperature sensor and operation of the thermal control system improve the thermal control appreciably. This control method was used during 24 VCE tests with water as a test liquid. The average temperature rise in these tests was 1.5°C during the transient stage and 0.2°C during steady state operation.


    Author Information:

    Meged, Y
    Haifa 34555,


    Stock #: JTE12650

    ISSN: 0090-3973

    DOI: 10.1520/JTE12650

    ASTM International is a member of CrossRef.

    Author
    Title Thermal Control of the Test Liquid in Vibratory Cavitation Erosion Tests
    Symposium , 0000-00-00
    Committee D09