Volume 29, Issue 4 (July 2001)

    Determination and Calculational Minimization of the Nonlinearity of a Vis-NIR-Spectrophotometer Using the Double Aperture Method

    (Received 28 January 2000; accepted 29 January 2001)

    CODEN: JTEOAD

      Format Pages Price  
    PDF Version 6 $25   ADD TO CART


    Abstract

    The determination of the nonlinearity of spectrophotometers is essential for evaluating instrument performance. Nonlinearity is cause mainly by the detector system. Indirect methods, e.g., using calibrated filters or standard solutions, to determine these small deviations are not reliable for high-accuracy measurements. Typical uncertainties are in the order of ± 0.003 Abs. (0.5 to 1% of stated transmittance). The double aperture method is based on the principle of light addition. It is a direct method that requires no secondary transmission or concentration measurements. With the double aperture mask, two light beams are produced that can be measured separately or additively. In an ideal linear system, the sum of the separate readings is equal to the additive reading. The deviation in a real system represents a degree of nonlinearity. Through simple mathematical considerations a correction term can be found.

    Using the double aperture method, the instrumental bias can be determined directly. While the measured systematic instrument error does not comprise additional uncertainties associated with the measurement of real samples, a calculational method to minimize instrument nonlinearity is shown.


    Author Information:

    Ossadnik, S
    Dipl.-Chem, assistant, TU Clausthal, Institute of Inorganic and Analytical Chemistry, Clausthal-Zellerfeld,

    Naundorf, G
    PerkinElmer (Switzerland) AG, Hünenberg,

    Schwedt, G
    Professor, TU Clausthal, Institute of Inorganic and Analytical Chemistry, Clausthal-Zellerfeld,


    Stock #: JTE12270J

    ISSN: 0090-3973

    DOI: 10.1520/JTE12270J

    ASTM International is a member of CrossRef.

    Author
    Title Determination and Calculational Minimization of the Nonlinearity of a Vis-NIR-Spectrophotometer Using the Double Aperture Method
    Symposium , 0000-00-00
    Committee E12