Reference Verification Fluids for Flash Point Determination

Montemayor, RG
Imperial Oil Ltd., Sarnia, Ontario

Rogerson, JE
Cedar Lane Farm, Maineville, OH

Colbert, JC
National Institute of Standards and Technology (NIST), Gaithersburg, MD

Schiller, SB
NIST, Gaithersburg, MD

(Received 17 August 1998; accepted 18 August 1999)

Abstract

The flash point of a liquid material is used to classify whether the material is flammable or combustible. 1,4-dimethylbenzene (p-xylene) has been used historically to verify the proper operation of different flash point test equipment. Because of the toxicity and low flash point value of 1,4-dimethylbenzene, an interlaboratory study was conducted to determine alternate flash point reference standards. This study reports the flash points of n-decane, n-undecane, n-tetradecane, and n-hexadecane using various flash point test methods. The flash points of these materials were found to be testmethod dependent because of the inherent differences in the various flash point test methods. The reported reference flash point values of these materials are now used to verify the proper operation of flash point test equipment in place of 1,4-dimethylbenzene.



Keywords:
flash point, verification fluids, n-decane, n-undecane, n-tetradecane, n-hexadecane, p-xylene

Paper ID: JTE12171J
DOI: 10.1520/JTE12171J