Volume 20, Issue 4 (July 1992)

    An Evaluation of Strain Measuring Devices for Ceramic Composites

    (Received 2 August 1991; accepted 30 October 1991)

    CODEN: JTEOAD

      Format Pages Price  
    PDF 11 $25   ADD TO CART


    Abstract

    A series of tensile tests were conducted on SiC RBSN composites using different methods of strain measurement. The tests were used to find the optimum strain sensing device for use with continuous fiber reinforced ceramic matrix composites in ambient and high temperature environments. Bonded resistance strain gages were found to offer excellent performance for room temperature tests. The clip-on gage offers the same performance but significantly less time is required for mounting it to the specimen. Low contact force extensometers track the strain with acceptable results at high specimen temperatures. Silicon carbide rods with knife edges are preferred. The edges must be kept sharp. The strain measuring devices should be mounted on the flat side of the specimen. This is in contrast to mounting on the rough thickness side.


    Author Information:

    Gyekenyesi, JZ
    Full-Time Research Associate for Cleveland State University at NASA Lewis Research Center, Cleveland, OH

    Bartolotta, PA
    Senior Research Engineer, NASA Lewis Research Center, Cleveland, OH


    Stock #: JTE11724J

    ISSN: 0090-3973

    DOI: 10.1520/JTE11724J