Digital Library / Journals / Journal of Testing and Evaluation (JTE) / Citation Page


Volume 20, Issue 3 (May 1992)

ISSN: 0090-3973
Page Count: 4


Comparison of Materials for Staple Pull-Out Tests
Thomas, AMC
The Research and Teaching Centre, Royal Orthopaedic Hospital,

Pynsent, PB
The Research and Teaching Centre, Royal Orthopaedic Hospital,

McCombe, PF
The Research and Teaching Centre, Royal Orthopaedic Hospital,

Abstract

Maximum pull-out forces of 16 mm bridge width staples of 10, 15, 20, and 25 mm leg length were tested in cadaver femoral metaphyseal bone. Pull-out force was related to leg length, but there was a very wide standard deviation. A plastic composite bone and beech hardwood were tested as substitutes and found to be unsuitable. A balsa wood composite was found to provide more reproducible results and is recommended for staple testing.



Keywords:
staple, pull-out force, retention force, powered stapling

Paper ID: JTE11717J
DOI: 10.1520/JTE11717J
ASTM International is a member of CrossRef.

Author Title Comparison of Materials for Staple Pull-Out Tests Symposium , 0000-00-00 Committee F04