Uncertainty Analyses on Reference Values of Charpy Impact Test Specimens

    Volume 2, Issue 9 (October 2005)

    ISSN: 1546-962X

    CODEN: JAIOAD

    Page Count: 13


    Takagi, S
    National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Ibaraki,

    Yamaguchi, Y
    National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Ibaraki,

    Usuda, T
    National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Ibaraki,

    (Received 23 November 2004; accepted 23 February 2005)

    Abstract

    The uncertainties of reference values of specimens, which are used for the verification of the secondary standard Charpy impact test machine, are investigated. To evaluate the uncertainty of the reference values of specimens, the results of the direct verification of the machine and the repeatability of measurement are considered. The uncertainty components estimated from the direct verification are calculated multiplying the uncertainty of each item and corresponding sensitivity coefficient. The repeatability of measurement is estimated from the results of the routine check data of the standard machine. By combining these components, the overall uncertainties of reference specimens are presented at 30, 100, and 160 J levels of absorbed energy.


    Paper ID: JAI12871

    DOI: 10.1520/JAI12871

    ASTM International is a member of CrossRef.

    Author
    Title Uncertainty Analyses on Reference Values of Charpy Impact Test Specimens
    Symposium Pendulum Impact Machines: Procedures and Specimens, 2004-11-08
    Committee E28