ISSN: 1546-962X
Published Online: 19
July 2006
Page Count: 8
Verification of Strain Level Calculations in Nitinol Fatigue Resistance Predictions
Pike, K
Abbott Vascular,
CA
(Received 11 October 2005; accepted 14 June 2006)
Abstract
A method to confirm FEA strain level predictions was developed for small nitinol-compressive specimens by taking micron magnitude measurements from SEM images. The technique was used to confirm predictions of an earlier report of fatigue resistance.
Keywords:
fatigue, nitinol, compressive, SEM, stent, FEA (Finite Element Analysis)
Paper ID: JAI100391
DOI: 10.1520/JAI100391
ASTM International is a member of CrossRef.
Author
Title Verification of Strain Level Calculations in Nitinol Fatigue Resistance Predictions
Symposium Fatigue and Fracture of Medical Metallic Materials and Devices, 2005-11-11
Committee E08