ISSN: 1546-962X
Page Count: 4
Gravimetric Analysis of the Nonvolatile Residue from an Evaporated Droplet, Using the Quartz Crystal Microbalance/Heat Conduction Calorimeter
Smith, AL
Masscal Corporation,
MA
Abstract
A quartz crystal microbalance/heat conduction calorimeter (QCM/HCC) has been used to determine the nonvolatile residue at the parts-per-million level in 10 µL droplets of methanol and a 45 µL droplet of water. The QCM was used to measure the mass of the nonvolatile residue, and the calorimeter was used to determine the mass of the solvent by the heat required to evaporate the droplet.
Keywords:
gravimetric analysis, nonvolatile residue, quartz crystal microbalance, extractable residue
Paper ID: JAI13894
DOI: 10.1520/JAI13894
ASTM International is a member of CrossRef.
Author
Title Gravimetric Analysis of the Nonvolatile Residue from an Evaporated Droplet, Using the Quartz Crystal Microbalance/Heat Conduction Calorimeter
Symposium Techniques in Thermal Analysis:Hyphenated Techniques, Thermal Analysis of the Surface, and Fast Rate, 2004-05-25
Committee E37