Digital Library / Journals / Journal of ASTM International (JAI) / Citation Page


Volume 2, Issue 9 (October 2005)

ISSN: 1546-962X
Page Count: 10


Development and Evaluation of a TG/DTA/Raman System
Collins, WJ
Senior Researcher, Procter & Gamble Pharmaceuticals, NY

DuBois, C
Senior Scientist, Procter & Gamble Pharmaceuticals, NY

Thomas Cambron, R
Principle Scientist, Procter & Gamble Pharmaceuticals, NY

Redman-Furey, NL
Senior Researcher, Procter & Gamble Pharmaceuticals, NY

Bigalow Kern, AS

Abstract

Modification of a commercially available TG/DTA instrument enabled development of a working TG/DTA/Raman system. The TG/DTA was modified by replacing a portion of the furnace wall above the sample and reference holders with a double layer insulated quartz window. The quartz window provided ready access for Raman spectroscopy, enabling simultaneous collection of TGA, DTA, and Raman data during thermal analytical experiments. Insertion of the quartz glass window did not impact TGA or DTA performance as demonstrated by the comparison of baseline drift and temperature calibration for modified and unmodified furnaces.



Keywords:
Thermal Analysis, TGA, TG/DTA, Raman, Raman Spectroscopy

Paper ID: JAI12789
DOI: 10.1520/JAI12789
ASTM International is a member of CrossRef.

Author Title Development and Evaluation of a TG/DTA/Raman System Symposium Techniques in Thermal Analysis:Hyphenated Techniques, Thermal Analysis of the Surface, and Fast Rate, 2004-05-25 Committee E37