Autopsy Artifact Created by the Revivant AutoPulse™ Resuscitation Device

    Volume 50, Issue 1 (January 2005)

    ISSN: 0022-1198

    CODEN: JFSOAD

    Published Online: 1 January 2005

    Page Count: 5


    Hart, AP
    Office of the Chief Medical Examiner, San Francisco, CA

    Stephens, BG
    Office of the Chief Medical Examiner, San Francisco, CA

    Hart, KR
    New College of California, San Francisco, CA

    Azar, VJ
    Office of the Chief Medical Examiner, San Francisco, CA

    (Received 10 April 2004; accepted 31 July 2004)

    Abstract

    In certain cases, the evaluation and correct identification of resuscitative artifacts is critical to the correct diagnosis and determination of the cause and manner of death. Resuscitative artifacts can resemble homicidal or accidental injury and thus possibly be misinterpreted. Occasionally, new technologies and/or medical procedures will create original and/or distinctive artifacts. In 2003, the San Francisco Fire Department emergency personnel began field-testing the Revivant AutoPulse™, an automated chest compression device. This device is currently being used in two other counties in the San Francisco Bay Area as well as regions of Florida, Virginia, and Ohio. We present three cases of resuscitative artifact that could be potentially confused with homicidal or accidental injury. These cases illustrate resuscitative artifacts, specifically lateral chest and horizontally oriented upper abdomen cutaneous abrasions created by this automated chest compression device.


    Paper ID: JFS2004155

    DOI: 10.1520/JFS2004155

    ASTM International
    is a member of CrossRef.

    Author
    Title Autopsy Artifact Created by the Revivant AutoPulse™ Resuscitation Device
    Symposium , 0000-00-00
    Committee E30