ASTM WK39462

    Work Item: ASTM WK39462 - Revision of D7094 - 04(2009) Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP) Tester

    (What is a Work Item?)

    Active Standard: D7094 - 12e1

    Developed by Subcommittee: D02.08 | Committee D02 | Contact Staff Manager



    1. Rationale

    When D7094 was first published, only instruments from one manufacturer were commercially available. Footnote 5 in the method references this sole source of supply. Since 2011, instruments from at least one additional manufacturer are available on the market. In Form and Style for ASTM Standards (Blue Book), ASTM states: F4.3.2 Reference to specific commercial sources of supply are permitted only when there is a sole source of supply. The new apparatus is technically equivalent to the original apparatus and conforms in every detail to the requirements of D7094 as described in the apparatus section and the Annex on Apparatus Specifications. A limited equivalency study was performed on two sites to demonstrate that the new apparatus is equivalent to the original one and meets the published precision. The results of this study were presented at the last ASTM D02 meeting. The study is described in the accompanying research report put up for administrative ballot. Subcommittee D02.08.0B decided to ballot the removal of the footnote 5. Footnote 7 in D7094 refers to the research report on the original ILS. It is proposed to also update footnote 7 with the information about which type of apparatus was used to develop the precision statement, in accordance with the ASTM D02 Facts for Members, Sec. 5.1.


    Keywords

    continuously closed cup; flammability; flash point; Closed-cup flash point methods; Flash point ;
    Citing ASTM Standards
    Back to Top
    Work Item Status

    Date Initiated:
    10-24-2012

    Technical Contact:
    Roland Aschauer

    Status:
    Draft Under Development