1. Scope
1.1 This practice covers the handling, testing, and procedure for using a lithium fluoride (LiF)-based photo-fluorescent film dosimetry system to measure absorbed dose (relative to water) in materials irradiated by photons or electrons. Other alkali halides that may also exhibit photofluorescence (for example, NaCl, NaF, and KCl) are not covered in this practice. Although various alkali halides have been used for dosimetry for years utilizing thermoluminescence, the use of photoluminescence is relatively new.
1.2 This practice applies to photo-fluorescent film dosimeters (referred hereafter as photo-fluorescent dosimeters) that can be used within part or all of the following ranges:
1.2.1 Absorbed dose range of 5 10-2 to 3 102 kGy (1-3).
1.2.2 Absorbed dose rate range of 0.3 to 2 10 4 Gy/s (2-5)).
1.2.3 Radiation energy range for photons of 0.05 to 10 MeV (2).
1.2.4 Radiation energy range for electrons of 0.1 to 10 MeV (2).
1.2.5 Radiation temperature range of -20 to +60°C (6,7).
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents
E170 Terminology Relating to Radiation Measurements and Dosimetry
E275 Practice for Describing and Measuring Performance of Ultraviolet, Visible, and Near Infrared Spectrophotometers
E925 Practice for the Periodic Calibration of Narrow Band-Pass Spectrophotometers
51204 Practice for Dosimetry in Gamma Irradiation Facilities for Food Processing
51261 Guide for Selection and Calibration of Dosimetry Systems for Radiation Processing
51431 Practice for Dosimetry in Electron and Bremsstrahlung Irradiation Facilities for Food Processing
51608 Practice for Dosimetry in an X-ray (Bremsstrahlung) Facility for Radiation Processing
51649 Practice for Dosimetry in an Electron Beam Facility for Radiation Processing at Energies between 300 keV and 25 MeV
51702 Practice for Dosimetry in a Gamma Irradiation Facility for Radiation Processing
51707 Guide for Estimating Uncertainties in Dosimetry for Radiation Processing
51818 Practice for Dosimetry in an Electron Beam Facility for Radiation Processing at Energies between 80 keV and 300 keV
51956 Practice for Thermoluminescence-Dosimetry (TLD) Systems for Radiation Processing
ICRU Report 14 Radiation Dosimetry: X-rays and Gamma rays with Maximum Photon Energies Between 0.6 and 50 MeV
ICRU Report 17 Radiation Dosimetry: X-rays Generated at Potentials of 5 to 150 kV
ICRU Report 34 The Dosimetry of Pulsed Radiation
ICRU Report 35 Radiation Dosimetry: Electron Beams with Energies Between 1 and 50 MeV
ICRU Report 60 Fundamental Quantities and Units for Ionizing Radiation
Index Terms
absorbed dose; calibration facility; dosimeter; dosimeter calibration; dosimetry system; electron beam; food irradiation; gamma ray; ionizing radiation; photo-fluorescent dosimeter; quality assurance; radiation processing; routine dosimeter; ICS Number Code 17.240
DOI: 10.1520/E2304-03